Flourite, also known as the mineral fluorspar, is the most valuable source of fluorine. The mineral’s purity — that is, calcium fluoride (CaF2) content — will establish its ultimate application in the aluminum, chemical, and steel sectors. The highest grade of fluorspar is used for producing hydrogen fluoride — a precursor to several major compounds including polymers and pharmaceuticals, for example, polytetrafluoroethylene (PTFE), commercially known as Teflon.
Therefore, it is vital to establish fluorspar’s fluorine content for both quality control and commercial purposes.
In traditional wet chemical techniques used for determining CaF2, calcium is measured by complexometric titrimetry (ASTM E815), and fluoride is determined by ion chromatography or ion selective electrode. However, all these methods involve skilled chemists, time-intensive operations, and require the use of potentially dangerous chemicals, which have to be subsequently disposed, thus collectively contributing to the cost of the analysis.
The MQC benchtop nuclear magnetic resonance (NMR) analyzer provides a rapid, simple, and non-destructive technique for measuring fluorine in fluorspar, which could be employed for routine analysis in a production setting without any need for specialist operator training and extra chemicals.
The analytical method is based on direct measurement of the nuclear magnetic resonance (NMR) signal of fluorine-19, which is known to have 100% natural abundance. It is believed that the quantified NMR response is linearly proportional to a material’s CaF2 content. The NMR signal, thus obtained, is normalized by the sample mass and the CaF2 content (weight-%) is subsequently computed using a suitable calibration curve.
The MQC benchtop NMR analyzer can be calibrated using just two samples, provided the reference values are known to be precise. Conversely, it is initially suggested to calibrate the instrument by 3-6 — if possible more — standards with known CaF2 contents uniformly spread across the range of interest. Since NMR is a comparative method, it cannot be more precise than the reference method against which it is being evaluated; analysis of more reference samples also reduces errors.
Figure 1. NMR calibration for calcium fluoride in fluorspar
The powder sample is transferred to a tared sample tube and filled up to a specified height and subsequently weighed before NMR analysis. It takes 80 seconds to measure each sample.
A comparison of the expected NMR values against the specified reference values is shown in Table 1. Deviations can be taken into consideration through an error in the reference values and also fluorine from sources other than CaF2 contributing to the NMR signal.
Table 1. NMR results predicted for calibration samples
In Table 2, the repeatability, or precision, of the NMR measurement of the same sample is shown to be excellent.
Table 2. Measurement repeatability of the same fluorspar sample
Oxford Instruments provides a package specifically customized to suit the measurement of fluorine in fluorspar.
This information has been sourced, reviewed and adapted from materials provided by Oxford Instruments Magnetic Resonance.
For more information on this source, please visit Oxford Instruments Magnetic Resonance.
Please use one of the following formats to cite this article in your essay, paper or report:
Oxford Instruments Magnetic Resonance. (2020, February 18). NMR Can Identify the Purity of the Mineral Fluorspar. AZoMining. Retrieved on August 20, 2022 from https://www.azomining.com/Article.aspx?ArticleID=1445.
Oxford Instruments Magnetic Resonance. "NMR Can Identify the Purity of the Mineral Fluorspar". AZoMining. 20 August 2022. <https://www.azomining.com/Article.aspx?ArticleID=1445>.
Oxford Instruments Magnetic Resonance. "NMR Can Identify the Purity of the Mineral Fluorspar". AZoMining. https://www.azomining.com/Article.aspx?ArticleID=1445. (accessed August 20, 2022).
Oxford Instruments Magnetic Resonance. 2020. NMR Can Identify the Purity of the Mineral Fluorspar. AZoMining, viewed 20 August 2022, https://www.azomining.com/Article.aspx?ArticleID=1445.
Do you have a review, update or anything you would like to add to this article?
This product profile outlines the Ramsey Series 30 Belt Scale System.
ZEISS Xradia 510 Versa x-ray microscope advances industry and science to address emerging research challenges.
Discover the PerGeos Digital Rock Software from Thermo Fisher Scientific.
AZoMining speaks to Philip Gross, CEO of Snow Lake Lithium, about the development of the world's first electric lithium mine. This is a particularly important development within mining and highlights the significance of accelerating green mining across the world.
AZoMining speaks with James Tour from Rice University about his team's exciting method that has the very real potential to recover valuable REE from three types of waste; electronic waste, coal fly ash and bauxite residue.
Craig Liddicoat from Flinders University speaks to AZoMining about his team's work on effective rehabilitation in post-mining ecosystems and whether it can set up a predictable trajectory of recovery. We also ask why ecosystem recovery is important in mining operations and the current limitations af
AZoMining.com - An AZoNetwork Site
Owned and operated by AZoNetwork, © 2000-2022